Ryan Nord, MD · Podcast

OSA Paper Podcast

Every day, one new paper on obstructive sleep apnea surgery, read in full and placed against a library of more than four hundred papers: hypoglossal nerve stimulation, palate surgery including barbed reposition pharyngoplasty (BRP), tongue base, skeletal and nasal procedures, patient selection and comparators. Each episode runs about fifteen minutes in four parts: the paper in two minutes, a teaching round, where it sits in the literature, and whether it should change practice.

Episodes

Newest first. Each title links to the paper; the player streams the episode.

November 17, 2026 · 13 min

The Association of Upper Airway Anatomy and Hypoglossal Nerve Stimulation Response in OSA

Phung C, Harkins T, Magaña LC, Dixit Y, Thuler E, Aggarwal M, et al. The Laryngoscope 2026

A prospective study from Dedhia's group at the University of Pennsylvania asked whether tongue size relative to the oral cavity, measured on computed tomography, predicts response to hypoglossal nerve stimulation for obstructive sleep apnea. In sixty-five analyzed patients, it was not shown to predict response, though the study was too small to rule out a modest effect, and exploratory signals for hyoid position and upper jaw width remain unconfirmed. A large tongue alone should not exclude a candidate, and the authors write that their findings do not support routine computed tomography for hypoglossal nerve stimulation evaluation outside of surgical-planning contexts.

Email

November 16, 2026 · 13 min

Impact of Bariatric Surgery on Hypoglossal Nerve Stimulation Outcomes

Kaki PC, Wesson T, Choi A, Shah S, Kaffenberger T, Soose R, et al. The Laryngoscope 2026

A multicenter retrospective study in The Laryngoscope asked whether hypoglossal nerve stimulation works as well for people with obstructive sleep apnea who have had bariatric surgery. Among 72 matched patients, those with prior bariatric surgery had a larger drop in the apnea-hypopnea index, while response rates and daytime sleepiness did not differ significantly. The finding is one to watch: prior bariatric surgery is no reason to hesitate on the implant, but a better response remains a hypothesis.

Email

November 15, 2026 · 13 min

Clinical Severity Staging System and Response to Hypoglossal Nerve Stimulation for Obstructive Sleep Apnea

Ji J, Wang HZ, Jackson RS, Lindburg MR, Kallogjeri D, Landsness EC, et al. JAMA Otolaryngology-Head & Neck Surgery 2026

A single-center study from Washington University in Saint Louis asked whether neck size, body mass index, sleep apnea severity and comorbidity could predict who responds to hypoglossal nerve stimulation for obstructive sleep apnea. Combined into a staged score, these factors tracked a steady fall in response, with fewer than four in ten responding in the worst stage, though the score separated responders only moderately and has not been validated. Surgeons should watch for external validation and use the score for counseling rather than to select or exclude patients.

Email

November 14, 2026 · 14 min

Comparison of Esketamine Combined With Dexmedetomidine Versus Propofol Combined With Midazolam on Drug-Induced Sleep Endoscopy - A Randomized Controlled Trial

Jiang Y, Wang Z, Wen H, Wen Z, Chen L, Wang C, Cheng Y, Zhou R, Yu W, Yan J. BMC Anesthesiology 2026

A randomized trial of fifty adults at Shanghai Ninth People's Hospital asked whether esketamine with dexmedetomidine works better than propofol with midazolam as sedation for drug-induced sleep endoscopy. Time to the target sleep stage did not differ significantly between the regimens, and the newer regimen's fewer rescues and desaturations were exploratory findings, with no airway findings reported. Propofol-based, depth-monitored endoscopy remains the standard for hypoglossal nerve stimulation candidacy and palate surgery planning.

Email

November 13, 2026 · 14 min

Characteristics, Risk Factors, and Long‐Term Outcomes of Dysphagia Following Modified Uvulopalatopharyngoplasty: A Mixed‐Methods Study

Yan Y, Shi R, Wang Q, Guo R, Yao H, Wang W, et al. Clinical Otolaryngology 2026

Researchers at the First Affiliated Hospital with Nanjing Medical University followed adults with obstructive sleep apnea for a year after a conservative modified uvulopalatopharyngoplasty with tonsillectomy, to describe swallowing problems after surgery and who is most at risk. Swallowing complaints were common soon after surgery, and about 9 percent of patients (11 of 120) still screened positive at one year; female sex, reflux disease, anxiety or depression, and daytime sleepiness were linked to early symptoms. The study offers useful counseling figures, but with no preoperative baseline and no comparison group, its risk factors are worth watching until they are validated.

Email

November 12, 2026 · 14 min

Surgery for Sleep Apnea - Long-Term Results

Martínez-Moreno M, Martínez Ruiz de Apodaca P, Matarredona-Quiles S, González-Turienzo E, Dinu-Pistolea R, Espinosa-Arnau B, Carrasco-Llatas M. Acta Otorrinolaringologica Espanola 2026

A single-center Spanish study from Carrasco-Llatas's group asked whether the benefit of modern, non-resective palate surgery for obstructive sleep apnea lasts. In patients treated mostly with barbed reposition pharyngoplasty, the average apnea-hypopnea index more than halved and stayed below baseline in those retested out to about seven years, though only a small minority were still being tested at the longest follow-up. The findings strengthen counseling about durability and the value of repeat off-treatment sleep testing, rather than changing who is offered surgery.

Email

November 11, 2026 · 14 min

SNORES Classification - Multicenter Prospective Validation for Obstructive Sleep Apnea

Eishal Arshad, Junaid Iqbal, Aisha Shafique. The Laryngoscope 2026

A prospective study in The Laryngoscope tested whether SNORES, a six-domain score built from sleep testing, nasal airflow, breathing effort, awake endoscopy and sleep position, predicts later care in adults with obstructive sleep apnea at three hospitals in Pakistan. The score ranked patients for later changes to non-surgical treatment far better than the apnea-hypopnea index alone, with an area under the curve of 0.82, but it mainly predicted clinicians' own decisions and has not been validated elsewhere. The score is worth watching as a structured way to describe and counsel patients, while drug-induced sleep endoscopy and procedure-specific selection remain in place.

Email

November 11, 2026 · 13 min

Comparing Transoral Robotic Surgery and Hypoglossal Nerve Stimulation Using Mean Disease Alleviation in OSA

Mandloi S, DeKloe J, Kaki P, Shah R, Estephan LE, Naimi B, et al. The Annals of Otology, Rhinology, and Laryngology 2026

A single-center retrospective study from Thomas Jefferson University compared robotic lingual tonsillectomy with Inspire hypoglossal nerve stimulation in patients who could not tolerate continuous positive airway pressure, using mean disease alleviation, a measure that combines apnea-hypopnea index improvement with nightly use. After matching, the two operations showed no statistically significant difference, but the robotic surgery group was small (32 patients), younger and selected for different airway anatomy. It supports choosing between the operations by airway anatomy, but it cannot show that they are equivalent.

Email

November 10, 2026 · 14 min

CPAP Treatment Persistence Following Nasal Procedures in Obstructive Sleep Apnea - A Real-World Claims-Based Analysis of 1.3 Million Patients

Malhotra A, Takashima M, Vahabzadeh-Hagh A, Yaremchuk K, Niklewski P. Journal of Medical Economics 2026

A study drawing on a very large United States insurance claims database asked whether nasal procedures performed after starting continuous positive airway pressure were linked to patients staying on the therapy. Patients who had an office-based temperature-controlled radiofrequency treatment of the nasal valve stopped therapy at a lower rate than those without nasal surgery (hazard ratio 0.62), with smaller associations for septoplasty and for turbinate or sinus surgery, and no statistically significant association for functional rhinoplasty alone. The finding is one to watch: it supports examining the nose before declaring positive airway pressure a failure, but the design cannot show that the procedure caused the difference.

Email

November 8, 2026 · 13 min

Challenging Barriers in OSA - Outcomes of Down Under Pharyngoplasty and Coblation Tongue Channelling

Nasserallah MV, Monna M, Edwards Z, Lum M, Tobin V, Mahanta VR, Diamantaras D, Shetty S, Mulder J, de Silva N. European Archives of Oto-Rhino-Laryngology 2026

A two-center Australian cohort from Peninsula Health asked whether adults with obstructive sleep apnea and so-called unfavorable throat anatomy can still benefit from a modified barbed reposition pharyngoplasty combined with coblation tongue channelling. Among patients who had a follow-up sleep study, about three in four met standard criteria for surgical success, including many with anatomy traditionally thought to predict failure. Because the study had no control group, short follow-up and incomplete sleep-study data, the finding is a promising signal to watch rather than proof.

Email

November 7, 2026 · 13 min

Hypoglossal Nerve Stimulation and the Incidence of Cardiovascular Disease

Kondamuri N, Hyman MJ, Cai Y, LoSavio PS, Skolarus TA. JAMA Otolaryngology-Head & Neck Surgery 22 -

A University of Chicago study in JAMA Otolaryngology–Head and Neck Surgery used a national commercial insurance database to ask whether hypoglossal nerve stimulation is linked to fewer new cardiovascular diagnoses in patients with obstructive sleep apnea who could not use positive airway pressure. With longer follow-up, implanted patients had about half the hazard of a new hypertension diagnosis, but early on they had more new hypertension diagnoses, and heart attack, stroke and heart failure were not significantly reduced. The finding is one to watch: it informs patient counseling but does not support offering the implant for cardiovascular protection.

Email

November 6, 2026 · 13 min

Low Arousal Threshold and Long-Term Adherence to Hypoglossal Nerve Stimulation

Kaki PC, Ravuri S, Patel SD, Kim J, Edwards BA, Vena D, et al. JAMA Otolaryngology-Head & Neck Surgery 2026

A two-center retrospective study of Inspire hypoglossal nerve stimulation patients asked whether a low respiratory arousal threshold, estimated from routine sleep study reports, predicts falling away from therapy. Early adherence was similar, but by six months low-threshold patients were far more likely to have lapsed, mostly by skipping whole nights. The take-home point: the signal supports closer counseling and follow-up for these patients, not a different decision about who is offered an implant.

Email

November 5, 2026 · 14 min

Adenotonsillectomy for Children With Mild Sleep-Disordered Breathing - Who Benefits?

Chervin RD, Gueye-Ndiaye S, Cen M, Mitchell RB, Liu C, Ibrahim S, et al. Sleep 2026

A secondary analysis of the Pediatric Adenotonsillectomy Trial for Snoring asked whether children who snore but show little or no obstructive sleep apnea on a sleep study benefit from early adenotonsillectomy, and whether any feature known before surgery predicts who benefits. Surgery beat watchful waiting on a combined measure of behavior, quality of life, sleepiness and blood pressure at one year, with one extra child meaningfully improved for every four treated, but almost no baseline feature predicted benefit apart from a possible signal from a high Mallampati score. The study supports counseling families that a near-normal sleep study need not by itself rule out surgery, while the Mallampati signal is one to watch until it is replicated.

Email

November 4, 2026 · 13 min

Phrenic Nerve Stimulation Enhances Upper Airway Patency During Drug-Induced Sleep Endoscopy in Obstructive Sleep Apnea

Yu JL, Gouldman KP, Kane BL, Manning KN, McEachran MC, Bourn D, Schneider H, Schwartz AR. Annals of the American Thoracic Society 2026

A 2026 study in the Annals of the American Thoracic Society asked whether stimulating the phrenic nerve through the skin, timed to different points in the breath, could improve upper airway flow in sedated patients with obstructive sleep apnea undergoing drug-induced sleep endoscopy. In 13 patients, a single burst at the end of expiration raised peak inspiratory flow by about 64 percent compared with unstimulated breaths, while bursts late in the breath did not help. Because the study measured no apnea-hypopnea index and no natural sleep, it is a mechanism finding to watch rather than a reason to change who is offered surgery.

Email

November 3, 2026 · 14 min

Effect of Novel Hypoglossal Stimulation on Airflow and Airway Collapsibility in OSA

Osman AM, Carney AS, Ooi EH, Joshi H, Toson B, Tran C, Hudson AL, Nguyen P, Gelekis N, Eastwood PR, Catcheside P, Eckert DJ. Chest 2026

A study in CHEST from Eckert's group at Flinders University asked whether a temporary electrode, placed through the skin under ultrasound guidance next to the hypoglossal nerve, could reopen the airway in sedated adults with obstructive sleep apnea during drug-induced sleep endoscopy. Stimulation restored airflow almost to the level reached on continuous positive airway pressure in thirteen of fourteen participants. With no natural sleep, apnea-hypopnea index or implantable device studied, the approach is one to watch rather than a reason to change who is offered hypoglossal nerve stimulation today.

Email

November 2, 2026 · 13 min

Physiologic Mechanisms of Response to Hypoglossal Nerve Stimulation

Magaña LC, Dixit Y, Saconi B, Keenan BT, Seay EG, Schwartz AR, et al. Sleep 2026

Researchers at the University of Pennsylvania asked whether airway physiology measured during drug-induced sleep endoscopy explains why some patients respond to hypoglossal nerve stimulation and others do not. In thirty patients with Inspire implants, the drop in airway opening pressure when the stimulator was switched on tracked the fall in the back-sleeping apnea-hypopnea index, with a strong correlation of 0.70, while measurements taken before stimulation did not separate responders from nonresponders. The verdict is watch: the finding helps explain response and frames the endoscopy workup of non-responders, but it does not yet provide a test for choosing patients before implant.

Email

November 1, 2026 · 14 min

Association Between Body Mass Index and Unilateral Hypoglossal Nerve Stimulation Efficacy Depends on the Pattern of Pharyngeal Collapse in Obstructive Sleep Apnea

Vena D, Kezirian EJ, Wellman A, Kent D, D’Agostino M, Monteiro JLGC, et al. American Journal of Respiratory and Critical Care Medicine 2026

A 2026 study in the American Journal of Respiratory and Critical Care Medicine asked whether body mass index affects the results of hypoglossal nerve stimulation differently depending on the pattern of airway collapse seen on drug-induced sleep endoscopy. In patients whose airway walls collapsed from the sides, higher body mass index was linked to substantially less reduction in the apnea-hypopnea index, while in front-to-back collapse the effect was small and not statistically significant, though the side-wall group held only sixty patients. What it adds: the finding supports counseling by collapse pattern rather than body mass index alone, but it needs independent confirmation.

Email

October 30, 2026 · 14 min

Anthropometric Measurements Inform Complete Concentric Collapse Status in Patients With Obstructive Sleep Apnea

Weiner JS, Chio EG, Dedhia RC, Huntley CT, Hutz MJ, Huyett P, et al. OTO Open 2026

A prospective multicenter study from Weiner's group in OTO Open asked whether body mass index, neck circumference and supine pharyngeal width could predict complete concentric collapse on drug-induced sleep endoscopy in candidates for hypoglossal nerve stimulation. A decision tree combining body mass index and neck circumference was right about 96 percent of the time when it called a patient clear, only modestly better than assuming no one had the finding. The finding is one to watch: sleep endoscopy remains part of candidacy screening, while these simple measurements offer useful numbers for counseling.

Email

October 29, 2026 · 14 min

Deep Learning-Based Automatic Scoring of Drug-Induced Sleep Endoscopy in Obstructive Sleep Apnea

Kim JY, Mun SJ, Baik YS, Lee YS, Kim YJ, Oh J, et al. npj Digital Medicine 2026

A five-hospital Korean study asked whether a deep learning model could grade drug-induced sleep endoscopy videos the way an expert panel does, scoring both the degree of airway obstruction and its primary cause. On videos from hospitals not used in training, the model matched expert grading of soft palate obstruction in 82.4 percent of 717 clips, with weaker performance lower in the airway, especially at the epiglottis. It is a tool to watch: it cannot grade concentric collapse and has not been linked to treatment outcomes, so it does not yet inform hypoglossal nerve stimulation candidacy.

Email

October 28, 2026 · 14 min

Subjective and objective measures of sleep-related function from the Cardiovascular Endpoints For Obstructive Sleep Apnea with Twelfth Cranial Nerve Stimulation (CARDIOSA-12) trial: Sleep-related functional outcomes in CARDIOSA-12

Tangutur A, Yu PK, Basner M, Keenan BT, Dedhia RC. Journal of Clinical Sleep Medicine 2026

A secondary analysis of the Cardiovascular Endpoints for Obstructive Sleep Apnea with Twelfth Cranial Nerve Stimulation (CARDIOSA-12) randomized crossover trial asked whether full versus partial hypoglossal nerve stimulation improves objective cognition and patient-reported sleep outcomes in obstructive sleep apnea. Sleepiness, insomnia, snoring and daily-function questionnaire scores improved on full stimulation, with the Epworth sleepiness score about 1.75 points lower, but reaction-time and processing-speed tests showed no statistically significant difference in the main analysis. The findings leave candidacy as it was but support counseling patients to expect feeling better, not measurably sharper thinking.

Email

October 27, 2026 · 13 min

Three-Year Outcomes of Proximal Hypoglossal Nerve Stimulation in OSA

Schwartz AR, Jacobowitz O, Mickelson SA, Miller MB, Oliven A, Certal V, et al. Chest 2026

Schwartz and colleagues, with Maurer as senior author, asked whether proximal hypoglossal nerve stimulation, a LivaNova device distinct from Inspire, keeps working at three years in the Targeted Hypoglossal Neurostimulation 3 trial. In the patients who completed a three-year sleep study, the median apnea-hypopnea index was roughly halved and held steady from year one, though about one in four patients did not finish follow-up and no three-year responder rate was reported. Take-home point: useful counseling evidence for patients asking about the other nerve stimulator, but not a reason to change Inspire candidacy criteria.

Email

October 27, 2026 · 14 min

Hypoglossal Nerve Stimulation and Hypoxic Burden in Patients With Obstructive Sleep Apnea: A Secondary Analysis of the STAR Trial

Xu J, Hajipour Z, Aloia MS, de Beeck SO, Vanderveken OM, Huyett P, et al. JAMA Otolaryngology-Head & Neck Surgery 2026

A secondary analysis of the Stimulation Therapy for Apnea Reduction trial asked whether hypoglossal nerve stimulation with a first-generation Inspire device lowers hypoxic burden, a measure of how deep and how long nighttime oxygen drops are. About seven in ten patients had their hypoxic burden cut by more than half at one year, though the study had no control group and measured no heart events. The finding is one to watch and adds a useful counseling point, but it does not change who is offered implantation.

Email

October 24, 2026 · 14 min

Proximal Hypoglossal Nerve Stimulation for Obstructive Sleep Apnea in the OSPREY Study: A Randomized Controlled Trial

Malhotra A, Schwartz AR, Lovett E, Juran N, Nguyen SA, Barrera JE, et al. Annals of Internal Medicine 2026

The OSPREY randomized trial tested whether an investigational proximal hypoglossal nerve stimulator, the LivaNova aura6000, works better than the same device implanted but switched off in adults with moderate to severe obstructive sleep apnea who could not tolerate positive airway pressure. At seven months, 58.2 percent of 67 treated patients met the response criteria, compared with 13.5 percent of 37 controls, with no serious device-related adverse events. The trial gives surgeons a randomized benchmark for counseling patients about an investigational alternative, but it is not Inspire evidence and does not address complete concentric collapse, so it is one to watch.

Email

October 21, 2026 · 14 min

A Randomized Trial of Nefopam for Postoperative Pain Management in Obstructive Sleep Apnea Patients Undergoing Uvulopalatopharyngoplasty

Chumpathong S, Mangmeesri P, Panyamee S, Banhiran W, Uataya M. European Archives of Oto-Rhino-Laryngology 2026

Chumpathong and colleagues at Siriraj Hospital in Bangkok ran a small randomized, placebo-controlled trial asking whether nefopam, a non-opioid pain drug not approved in the United States, reduces morphine use in the first day after uvulopalatopharyngoplasty for obstructive sleep apnea. Added to a strong multimodal pain regimen, it did not: both groups needed only a few milligrams of morphine, with no statistically significant difference. Because the trial was underpowered, it supports not adding nefopam routinely while leaving a modest effect unexcluded.

Email

October 21, 2026 · 13 min

Systematic Review on the Safety of Outpatient Upper Airway Surgery for Obstructive Sleep Apnea Patients in Ambulatory Surgical Centers

Asghar M, Pang K, Asghar M, Rotenberg B. Journal of Otolaryngology - Head & Neck Surgery 2025

A 2025 systematic review from Rotenberg's group asked whether adults with obstructive sleep apnea can safely have nasal and palate surgery in ambulatory surgical centers and go home the same day. Across mostly older, hospital-based studies, unplanned admission for medical reasons averaged 5.4 percent across the four studies that reported it, above the authors' own 3 percent benchmark, and serious events were rare and came within the first few hours. The take-home point: the review maps early risk usefully but contains almost no true surgery-center data, so it does not yet justify moving pharyngoplasty patients out of overnight observation.

Email

October 20, 2026 · 14 min

Association Between Perioperative Celecoxib, Acetaminophen, and Opioid Consumption After Palate Surgery for Obstructive Sleep Apnea

Badash I, Wong S, Biju K, Hur K, Commesso E, Kezirian EJ. Otolaryngology-Head and Neck Surgery 2025

A single-surgeon retrospective study by Badash and colleagues, with Kezirian as senior author, asked whether celecoxib or acetaminophen given around palate surgery for obstructive sleep apnea was associated with less opioid use in the first day after surgery. Higher celecoxib doses were associated with modestly lower opioid use, while acetaminophen showed no significant association, and 205 of 210 patients still received opioids. Because celecoxib became routine partway through the study period, the finding is one to watch for confirmation in a randomized trial.

Email

October 18, 2026 · 13 min

Predictive Factors for Surgical Outcomes and Symptom Persistence After Uvulopalatopharyngoplasty in Patients With Obstructive Sleep Apnea

Ling X, Zheng J, Liu Q, Shi C, Fan C. Head & Face Medicine 2026

A single-center study from Wujin Hospital in China asked which preoperative features predict success after a modified uvulopalatopharyngoplasty for obstructive sleep apnea. About two in three patients met the success definition at short-term follow-up, and a lower Friedman stage, lower body mass index and higher lowest oxygen saturation were linked to success. The proposed prediction model has not been validated outside the patients who built it, so it is one to watch.

Email

October 17, 2026 · 14 min

Uvularis Muscle Strengthening Advancement in Palatopharyngeal Collapse in Selected Cases of Snoring and Obstructive Sleep Apnea: A Randomized Trial

Askar SM, Awad AM, Abou-Sharkh AA. European Archives of Oto-Rhino-Laryngology 2026

A randomized trial from Zagazig University in Egypt asked whether adding a uvularis muscle advancement stitch to modified anterior palatoplasty improves outcomes in young adults with mild or early moderate obstructive sleep apnea and single-level, front-to-back palatal collapse. In 34 patients, the added stitch did not significantly lower the apnea-hypopnea index compared with palatoplasty alone, although bed-partner-rated snoring improved more. The snoring benefit is one to watch, pending a blinded, registered trial.

Email

October 16, 2026 · 14 min

Uvulopalatopharyngoplasty Versus Expansion Sphincter Pharyngoplasty - A Single Centre Experience

Steinbichler TB, Bender B, Hartl R, Strasser V, Sontheimer D, Buricic S, et al. Clocks & Sleep 2025

A retrospective single-center study from the Medical University of Innsbruck compared expansion sphincter pharyngoplasty with a muscle-sparing uvulopalatopharyngoplasty in patients with snoring or obstructive sleep apnea, with the operation chosen by drug-induced sedation endoscopy findings. Expansion sphincter pharyngoplasty significantly lowered the apnea-hypopnea index, to about 13 events per hour, while the uvulopalatopharyngoplasty group's fall was not significant; expansion sphincter pharyngoplasty also brought more minor bleeding and more strong pain medication. The study adds independent support for palate operations that reshape the lateral pharyngeal wall and gives frank morbidity numbers for counseling, but it cannot prove superiority and does not test barbed reposition pharyngoplasty.

Email

October 15, 2026 · 14 min

Auditory Perceptions Associated With Hypoglossal Nerve Stimulation for Obstructive Sleep Apnea

Belilos A, Adibi I, Tucker A, Saeedi A, Calder A, Lee L, Nord R. Laryngoscope Investigative Otolaryngology 2026

Researchers at Virginia Commonwealth University surveyed Inspire hypoglossal nerve stimulation recipients by telephone to ask whether they hear new sounds when the device stimulates. About one in six reported a new buzzing or humming sound during stimulation, mostly mild, and none had stopped using the device; preexisting tinnitus was not a statistically significant predictor in this small sample. The finding changes how patients are counseled and questioned about sound at follow-up, not who is offered the implant.

Email

October 15, 2026 · 14 min

Preservation Palatopharyngoplasty for Obstructive Sleep Apnea: Predictors and Clinical Outcomes

Abdelwahab M, Alhussien A, Kassir MF, Butehorn H, Munhall C, Nguyen SA. Otolaryngology–Head and Neck Surgery 2026

A single-surgeon retrospective series from the Medical University of South Carolina asked how adults with moderate to severe obstructive sleep apnea fare after preservation palatopharyngoplasty, and which baseline features predict a larger drop in the apnea-hypopnea index. The apnea-hypopnea index was cut roughly in half, with larger tonsils and higher baseline severity predicting bigger drops, though only a minority of patients were cured. The take-home point: the paper sharpens counseling for patients with large tonsils who fall outside Inspire candidacy, but it needs controlled, longer-term confirmation.

Email

October 14, 2026 · 14 min

Post-Operative Radiological Evaluation of a Bilateral Hypoglossal Nerve Stimulation System

Kilgue A, Gehl H-B, Abolmaali N, Scholtz L-U, Scherl C, Todt I. Frontiers in Surgery 2026

A single-center German case series of eleven patients asked what the first postoperative image should be after implanting the Genio bilateral hypoglossal nerve stimulator. Plain x-rays showed the hardware cleanly in every patient, while every computed tomography scan carried metal artifact, and the paper reported no clinical outcomes linking paddle position to function. The take-home point: plain films make a sensible day-one baseline, and repeat imaging over time is needed to show whether the paddles migrate.

Email

October 14, 2026 · 14 min

Exploring the Role of Drug-Induced Sleep Endoscopy in Optimizing Upper Airway Stimulation Therapy for Non-Responders

Kant E, Hardeman JA, Stokroos RJ, Copper MP. Sleep and Breathing 2025

Copper's group at Sint Antonius Hospital in the Netherlands described how they use drug-induced sleep endoscopy with the stimulator switched on in patients who still fail Inspire hypoglossal nerve stimulation after standard titration, and what happened to those patients afterwards. In a small retrospective series without a control group, patients who had settings changes or added therapy and returned for follow-up saw their apnea-hypopnea index fall by about 18 events per hour. Dr. Nord considers the procedure a good tool for the tolerant non-responder once awake endoscopy and reprogramming have failed, while noting that the series cannot show the endoscopy itself caused the improvement.

Email

October 14, 2026 · 14 min

GLP-1 receptor agonists for obstructive sleep apnea: a meta-analysis of randomized controlled trials

Carvalhal G, Mora MMR, Peralta-Jiménez GA, Ayasa L, Barrera V, Advani K, et al. Sleep and Breathing 2026

A meta-analysis in Sleep and Breathing pooled randomized trials of liraglutide and tirzepatide in adults with moderate-to-severe obstructive sleep apnea. The drugs lowered the apnea-hypopnea index by about ten events per hour more than control, but the effect depended heavily on the drug, and pooled oxygen measures did not improve significantly. The paper does not change who is offered surgery, but it sharpens counseling: quote drug-specific numbers, re-test after weight loss, and, as in Dr. Nord's practice, treat weight-loss drugs and Inspire as complementary parts of treatment, though no trial has yet tested them together.

Email

October 14, 2026 · 14 min

Edema and Subjective Discomfort After Barbed Stayed Bridge Pharyngoplasty (BSBP)

Pace A, Iannella G, Maniaci A, Cocuzza S, Moffa A, Alunni Fegatelli D, et al. Journal of Clinical Medicine 2025

Researchers at Sapienza University of Rome followed a small, carefully selected group of patients with moderate obstructive sleep apnea after barbed stayed bridge pharyngoplasty, a muscle-sparing barbed palate operation, to measure early pain, swelling and discomfort. Pain stayed low, averaging about three out of ten on the first day and near zero by one month, with no bleeding, but without a comparison group the study cannot show the technique outperforms barbed reposition pharyngoplasty. Dr. Nord's take-home point: a promising early-recovery signal for one technique in thirty selected patients, still awaiting a head-to-head comparison with barbed reposition pharyngoplasty.

Email

October 14, 2026 · 14 min

Hypoxic Burden Reduction After Lateral Pharyngoplasty in Patients With Obstructive Sleep Apnea - Beyond the Apnea-Hypopnea Index

Vieira ALP, Duarte BB, Almeida AR, Greters ME, Leandro-Merhi VA, de Aquino JLB. Sleep and Breathing 2026

A single-center study from the Pontifical Catholic University of Campinas in Brazil asked whether lateral pharyngoplasty lowers hypoxic burden, a measure of nighttime oxygen loss, and not just the apnea-hypopnea index. In the 27 patients with hypoxic burden measured before and after surgery, median hypoxic burden fell to about a third of baseline. It also fell significantly in patients who failed by apnea-hypopnea index criteria, but the study had no control group and measured no cardiovascular outcomes. Verdict: watch — it supports reporting hypoxic burden alongside the apnea-hypopnea index after palate surgery, without changing who is offered surgery.

Email

October 11, 2026 · 15 min

Ultrasonographic Tongue Base Motion Does Not Correlate With Hypoglossal Nerve Stimulation Outcomes

Tschopp S, Janjic V, Caversaccio M, Borner U, Tschopp K. Laryngoscope Investigative Otolaryngology 2026

A two-center Swiss study asked whether the way the tongue base moves on awake ultrasound during hypoglossal nerve stimulation predicts how much a patient's obstructive sleep apnea improves. In 64 patients with unilateral Inspire implants, no ultrasound measure of tongue movement size, symmetry or shape was significantly associated with the drop in the apnea-hypopnea index, though the confidence intervals were wide. The call is no change in practice, with ultrasound worth watching as a research tool.

Email

October 7, 2026 · 23 min

Oropharyngeal Lateral Wall Collapse on Drug-Induced Sleep Endoscopy as a Predictor of Hypoglossal Nerve Stimulation Response in Obstructive Sleep Apnoea-A Systematic Review

McBrinn S, Skan O, Iacovidou A. Diagnostics 2026

A systematic review of eight Inspire cohorts, 1,051 patients in all, asked whether complete collapse of the lateral pharyngeal walls on sleep endoscopy predicts a poor response to hypoglossal nerve stimulation. It lowers the response rate by roughly 15 to 20 percentage points, yet about two of three such patients still respond, so it is a counseling point rather than an exclusion. Call: watch.

Email

October 5, 2026 · 17 min

Consensus Statements Among European Sleep Surgery Experts on Tongue, Hypopharynx, and Supraglottis Associated With Snoring and Obstructive Sleep Apnea - Part 1 - Evaluation and Decision Making

Olszewska E, De Vito A, Baptista P, Delakorda M, Heiser C, Cheong RCT, et al. Journal of Clinical Medicine 2025

A European expert panel reached consensus on how to evaluate the tongue base, hypopharynx and supraglottis in snoring and sleep apnea, and how those findings should steer treatment. The statements support staging palate surgery before hypoglossal nerve stimulation when large tonsils coexist with tongue base collapse, and repeating sleep endoscopy before any revision surgery. Call: watch; the workup stays as it is.

Email

Educational material for residents and clinicians. Episodes summarize published papers for teaching and discussion; they are not medical advice and do not replace the papers themselves. Site privacy notice.

Disclosures

Ryan Nord, MD, is a consultant to AvivoMed. He previously consulted for Inspire Medical Systems (ended 2025) and Medtronic, and is a clinical investigator for Nyxoah, the maker of Genio. This is his personal educational website. No manufacturer funds, reviews or sponsors the podcast or this page; industry-funded studies are identified when an episode cites them. See the full disclosures for his industry relationships and how to report a content correction.

Try “success rate”, “BMI”, “Medicare”, “battery”, or “concentric collapse”.